目 录
一 概述····················································································…… (1)
二 技术要求·................................................................................... (1)
三 检定条件·................................................................................... (1)
四 检定方法·................................................................................... (2)
五 检定结果的处理·.......................................................................... (4)
附录1 流速允许误差公式的说明···················································…… (5)
附录2 检定记录········································································…… (6)
附 录 3 检定证书背 面格 式 ················,···········································…… (7)
总悬浮颗粒物采样器检定规程
本检定规程适用于新制造、使用中和修理后的总悬浮颗粒物采样器 (以下简称为采
样器)的检定。
一 概 述
总悬浮颗粒物采样器用于测定空气中总悬浮颗粒物的质量浓度。其基本原理是:使
一定体积的空气通过已知质量的滤膜,悬浮于空气中的颗粒物被阻留在滤膜上,根据滤
膜增加的质量和通过滤膜空气的体积,确定空气中总悬浮颗粒物的质量浓度。采样器主
要由采样头、抽气动力装置、流量控制装置、定时装置等部分组成。
二 技 术 要 求
1 绝缘电阻:不小于20 Mn
2 外观
2.1 面板上应标明:采样器的名称、型号、编号、制造单位、出厂日期和制造计量器
具许可证标志与编号等。
2.2 部件应齐全,运转应正常,显示应清楚。
2.3 应无影响计量性能的损伤。
3 进气口宽度的允许误差
3.1 大流量:不超过士2.5%0
3.2 中流量:不超过士2%0
4 系统漏气量:不大于1%0
5 采样时间允许误差:不超过士0.5%0
6 流速允许误差:不超过18%0