目 录
一 概述 ·............................................................................................... (1)
二 技术要求 ·......................................................................................... (1)
三 检定条件 ················。······。···········································、··················一 (2)
四 检定项目和检定方法 ·…:··.................................................................... (2)
五 检定结果处理和检定周期 ····。·················。················。。·。。····················一 (4)
附录1检定用试剂·................................................................................... (g)
附录2检定记录格式·································································。·········……(6)
附录3 检定证书和检定结果通知书 (背面)格式········。······························…… (g)
附录4线性回归法求工作曲线·······························································……(9)
附录5 检定用高强度空心阴极灯稳定性 (暂行)检测方法 ························…… (10)
非色散原子荧光光度计检定规程
本规程适用于新制造、使用中和修理后的用空心阴极灯作光源的单、双道非色散氢
化物原子荧光光度计 (以下简称仪器)的检定。
一 概 述
该仪器可用于易形成氢化物的元素,易形成气态组份的元素和易还原成原子蒸气的
元素的测量。
该仪器是用由光源发出的特征辐射光照射在被测元素的原子蒸气上,基态原子被激
发到高能级,当以辐射去活化返回基态时,辐射出荧光的原理,来进行定量分析的。
其荧光强度与元素的浓度存在以下关系:
If=ol0(1一。一‘LN)
式中:介— 原子荧光强度;
45— 原子荧光量子效率;
Io— 光源辐射强度;
Kx— 在波长久时的峰值吸收系数;
L— 吸收光程;
N— 单位长度内基态原子数。
对于给定的元素来说,当光源的波长和强度固定,吸收光程固定,原子化条件一定,
在元素浓度较低时,荧光强度与荧光物质的质量浓度P有如下简单的关系 ((a为常数):
if=ap