目 录
技术要求 ····················································································…… (1)
一三 一
检定条件 ····················································································…… (1)
检定方法 ·············································……!’’······························…… (2)
四 检定结果处理和检定周期·························································,·····……(2)
附录 荧光亮度检测仪检定系统····························································……(4)
荧光亮度检测仪检定规程
本规程适用于新制造、使用中、修理后的指针式和数字式荧光亮度检测仪的检定。
荧光亮度检测仪是测量荧光渗透液受紫外辐射激发后其亮度特性的仪器,是通过测
量光照度来实现的,仪器包括光电探测器、指示仪表、壳体、电源盒和荧光板。
一 技 术 要 求
1荧光亮度检测仪光电探测器 [包括接收器、V(劝 滤光片、余弦修正器]的余弦修正
器表面应清洁无损,不得有斑点等痕迹。指针式仪表表面刻线清晰,指针不得有卡针、
弯曲、变形等缺陷。数字式仪表表面应清洁、显示清楚、完整。荧光板保持干燥,塑料
膜表面清洁,不许有发霉现象。电源盒应正常工作。
2 荧光亮度检测仪的光电探测器必须有 V(劝 滤光器,使它的光谱响应与国际照明委
员会CIE明视觉光谱光效率相匹配,匹配误差f -<5%.
3 荧光亮度检测仪的测量范围:光照度为(0一500) lx;辐照度为(0一2.8)nW/crri .
荧光亮度检测仪的测量标准不确定度:光照度为4% (501x一300川;辐照度为11%
(0.3mW/cmz一1.7mW/cm2).
荧光板的年稳定度为2%.
二 检 定 条 件
4 荧光亮度检测仪检定装置主要包括:紫外辐射源、白炽辐射源、标准荧光亮度检测
仪、标准荧光板、检定光照度的白色相纸、固定底板、调节支架和中性减光器。
4.1紫外辐射源为125W的高压黑光灯,辐射中心波长为J = 365mn,黑光灯泡壳磨毛,
垂直向下安装,辐射均匀,有黑色灯罩保护。
4.2 白炽辐射源为100W磨砂白炽灯泡,垂直向下安装,辐射均匀,有黑色灯罩保护。
4.3 标准荧光亮度检测仪由指针式表头或数字式表头、光电探测器、壳体、电源盒、标
准荧光板组成。
4.4 标准荧光板由特定的荧光粉加粘接剂制作而成,在一定强度的紫外辐射照射下所激
发出的荧光波长和强度稳定。
4.5 检定光照度所用的白色照相纸不得发黄。
4.6 中性减光器有足够的动态工作范围,使用它可使仪器指示到501x.
5 供电设备
220\交流稳压电源一台,稳压精度为*1%。出厂检定合格的125V高压汞灯镇流器一只。
6 紫外辐射源和白炽辐射源由固定支架支撑,并有3个自由度调节。工作平台稳定、牢
固 、平整 。