目 录
一 概述 ··························,·······································,····················…… (1)
二 技术要求 ,...................................................................................... ( 1 )
三 检定条件 ·...................................................................................... (2 )
四 检定项目和检定方法·....................................................................... (2 )
五 检定结果处理和检定周期 ·................................................................. ( 2 )
附录1 标准漫透射密度测量的几何条件 ·......................................····,一 (4)
附录2 标准视觉密度测量的光谱条件 ················································…… (5)
附录3 检定记录格式 ····,································································,,·… (6)
附录4 检定证书和检定结果通知书背面格式 ,·····················,··············,,,,,·(8)
器和显示器等组成。
二 技 术 要 求
1 外观要求
1.1 密度计应有下列标志:仪器名称、型号、编号、制造厂、出厂时间。
1.2 密度计应能平稳置于工作台上。各调节器、按键和开关的标志应清晰无误,并能
正常工作。各紧固件、电缆线的接插件均应紧密配合,接触良好。
1.3 密度计的显示器应显示清晰,数字显示不应有断缺线;指针显示的刻线、刻字等
应完整均匀,指示线宽度不应大于刻线宽度,并应与刻线平行。
1.4 密度计应配有专用的工作密度片,其密度片分光色片基和蓝色片基两种。
2 零点漂移
按第 10条规定的方法进行检定时,密度计零点漂移不超过士0.01.
3 重复性
按第 11条规定的方法进行检定时,密度计重复性应不大于0.01
4 稳定度
按第12条规定的方法进行检定时,密度计稳定度应不超过士0.010
5 示值误差