目 录
一 概述 ·······················································································…… (1)
二 技术要求 ·...................................................................................... ( 1 )
三 检定条件 ·...................................................................................... ( 3 )
四 检定项目和检定方法 ·....................................................................... ( 3 )
五 检定结果处理和检定周期 ·................................................................. ( 6 )
附录1 标准溶液及其配制方法 ·························································…… (7)
附录2 检定记录格式 ·····································································…… (9)
附录3 检定证书和检定结果通知书 (背面)格式 ·································…… (11)
一 概 述
仪器是根据物质分子对光的选择吸收特性和全差示分光光度法的朗伯一比尔
(Lambert一Beer)定律,在可见光区以相对测量法,对物质含量进行定量测定的分析仪
器。全差示分光光度法的朗伯一比尔定律的数学表达式为:
C=Cr+A '/aL
=Cr+(1/aL )1g( (i+1)/(£+:)〕 (1)
或: A"=aL(C一Cd
式中:A'— 物质的全差示吸光度;
C— 物质的浓度;
a— 物质的吸收系数;
:— 物质的透射比;
L— 液层的厚度;
Cr— 参比物质的浓度;
i— 全参示参数 (或称微电流参数)。
仪器主要由光源、单色器、样品室、检测器、全差示参数系统和指示器等组成。
仪器可直接进行普通分光光度法和高吸收差示法测量,并且可利用全差示参数系统
简单地进行低吸收差示法和最精密差示法的测量。
二 技 术 要 求
1 外观
1.1 仪器表面油漆和镀层不得有明显损伤和剥落现象。名称、型号、制造厂名、出厂
日期、仪器编号及制造许可证标志应清楚完整。
1.2 仪器应能平稳置于工作台上,各紧固件紧固良好,所有调节旋钮、开关和按键均
能正常工作,波长度盘无打滑现象,吸收池架拉杆能推拉自如、定位正确,各透光孔透
光量相同。
1.3 仪器指示器的刻线、刻字和读数标线应粗细均匀、清晰完整,指针宽度不大于刻
线宽度,并应与刻线平行。
1.4 仪器处于工作状态时,光源灯应无闪烁现象;全差示参数 :应能在全刻度范围内
连续可调;暗电流在灵敏度各档均可用零点调节器调至零点。波长置于580 nm处