目 录
一 概述 ··。··········,··,···································,··,··················,,········…… (1)
二 主要技术要求 ·................................................................................ ( 1 )
三 检定条件及检定器具 ·....................................................................... ( 3 )
四 检定项目及检定方法 ‘·······,·,,·,··。。···························,···········‘····一 (4)
五 检定结果处理及检定周期 ·................................................................. ( 6 )
附录1 电容式条干均匀度仪检定记录表 ·············································…… (7)
附录2 重量比较法 ,,··.......................................·‘························…… (10)
附录3 检测器输出端测量点 ····························································…… (10)
附录4 CV值调整方法 ··································································…… (11)
电容式条干均匀度仪检定规程
本规程适用于新制造、使用中及修理后的国内外生产的电容式条干均匀度仪 (以下
简称条干仪)的检定。
一 概 述
条于仪用于棉、毛、麻、绢、化学短纤维纯纺或混纺纱条、生丝、化学纤维长丝等
线密度不匀的测定和分析。基本组成包括检测器、变异系数 (以下用符号CV表示)计
算部分、测试结果输出部分等。根据不同的使用需要还可增设波谱分析、疵点计数、偏
移率 (以下用符号DR表示)和平均值系数 (以下用符号AF表示)的计算、CV一片
段长度曲线及片段试验等扩展部分。检测器采用电容检测法将纱条的线密度转换成相应
的电信号,经处理后以不同的方式输出。
二 主 要 技 术 要 求
1 条干仪应标有计量器具标志、计量器具生产许可证编号及铭牌。铭牌上须标明型号、
规格、制造厂、出厂编号和出厂年月。
条干仪的检测电容器极板、不匀指示器、平均值和长度系数调节旋钮、纱线测试路
径中的机械零部件等,均不应有影响计量性能的机械损伤。
新制造的条干仪面板和机壳不应有明显的划痕、裂缝、变形等现象;表面涂 (镀)
层不应起泡、龟裂和脱落;金属零件不应有锈蚀和缺损。
2 条干仪的纱条支架应完整齐全并能正常工作,纱条支架的转动部分应灵活可靠,并
可调节和控制。
3 张力器应能有效控制纱条的张力。具有加捻装置的条干仪其捻度和张力调节功能均
应正常有效。
4 电气安全性
条干仪应具有良好的电气安全性,其电源部分的绝缘电阻应)2 M氏 保护接地必
须安全可靠。
5 基本状态
5.1 纱线应正确通过细纱检测器各槽,应与检测电容器极板平行且不摩擦极板。
5.2 罗拉表面无明显的沟槽 (使用中无要求)、弯曲及中凹,罗拉离合功能正常。
5.3 不匀指示器的机械零点指示准确,无试样零点的指示偏差不超过100%量程挡的
士5% (新表要求指针不离线)。
5.4 平均值和长度系数调节旋钮在零位的指示偏差不超过一个最小分度。
5.5 在量程为 “12.5%”挡时,将不匀指示器的指针调至0%处,转换到其他挡时,