一 概述 ·······················································································…… (1)
二 技术要求 ·...................................................................................... ( 1 )
三 检定条件 ··,,·,,···············,·,·,···············,,,,·,··············,·,·,··············,一 (3)
四 检定项目和检定方法 ........................................................................ ( 3 )
五 检定结果的处理 ·..............................................................:.............. ( 6 )
六 检定周期 ·...................................................................................... ( 7 )
附录1 检定证书 (背面)格式 ···················。···················,···········‘·····,··… (8)
附录2 检定结果通知书 (背面)格式 ·‘···,············‘·····,················…… (9)
附录3 检定记录格式 ····································································…… (10)
附录4 校准溶液的配制 ··································································…… (13)
离子色谱仪检定规程
本规程适用于新制造、使用中和修理后,带有抑制型电导检测器的实验室离子色谱
仪的检定。其它类型的离子色谱仪可参照本规程和出厂仪器的技术指标进行检定
概 述
典型的实验室离子色谱仪由输液系统、进样装置、色谱柱、检测器和数据处理装置
等几部分组成。图1是典型的离子色谱仪组成示意图。
图 1 离子色谱仪组成示意图
样品注人到系统进样阀后,随流动相 (即洗脱液)经过保护柱进到分离柱中。由于
样品中各组分离子对色谱柱固定相的亲合力不同,因而不同种离子被先后洗脱而进人到
抑制器中,在抑制器中除去 (或降低)洗脱液的本底电导,并增加待测离子的电导响应
值,最后进人电导池,并按先后次序得到各待测离子的电导率。该电导率在低浓度下与
待测离子的浓度成正比。以待测组分的保留时间定性,以峰高或峰面积定量进行样品分
析。
二 技 术 要 求
1 外观
1.1 仪器应具备下列标志:名称、型号、制造厂名、出厂日期、系列号 (或编号)等。
1.2 仪器的各功能部件 (量程、输出旋钮、按键、开关和指示灯等)均能正常工作,
各紧固件应无松动。
1.3 仪器应标明所使用的电源、电压和频率。电源线、信号电缆等插头、接头应与插
座紧 密配合 。